Cummins/Onan GTA38 / GTA50 — Operating the Teledyne Model 320P / 320P-D Oxygen Meter

Operation · Service tool / oxygen meter

For the Cummins/Onan GTA38 / GTA50, from the Cummins/Onan owner's manual.

Applies to: Cummins/Onan GTA38 engine, Cummins/Onan GTA50 engine

Before you start

  • Excess moisture or temperature in the sample will damage the O2 cell.

Procedure

  1. Read manufacturer's manuals for initial startup.
  2. Check battery condition before each use; recharge if low (overnight, 14 hours).
  3. Check filter material; change if wet or dirty.
  4. Calibrate with ambient air (20.9% O2); install flow-through adapter cap.
  5. Connect temperature reducing tubing to engine exhaust system.
  6. Coil tubing several times or make a dip to act as a water trap.
  7. Depress pump button to pull exhaust sample; draw no longer than 5 minutes.
  8. Clear the water trap every 10 minutes.
  9. After each 5 minutes sampling, purge with clean ambient air to near 20.9%.
  10. Turn off unit and reinstall cell saver cap when not in use.

Tools required

  • Teledyne Model 320P/320P-D oxygen meter Amazon
  • 3/8 in copper or stainless tubing Amazon

Tool links are Amazon searches (affiliate links); manufacturer special tools may only be available from a dealer.

Specs & values

  • Ambient air 20.9% O2

Source: Cummins/Onan owner's manual. Steps and values are as documented; always cross-check against the official manual and any later revisions for your exact serial number before servicing.